Buradasınız

TİDAL VOLÜM, SOLUNUM SAYISI VE BASINÇ DEĞERİNİN BULANIK KONTROLÜ

FUZZY CONTROL OF TIDAL VOLUME, RESPIRATION NUMBER AND PRESSURE VALUE

Journal Name:

Publication Year:

Abstract (2. Language): 
In this study, control of tidal volume, respiration number and pressure value which are arrived to patient at mechanical ventilator device which is used in intensive care units were performed with fuzzy logic controller. The aim of this system is to reduce workload of aneshesiologist. By calculating tidal volume, respiration number and pressure value, the error Pe(k) between reference pressure value (Pref) and pressure of gas given ill person (Phasta) and error change rate Pe(k) were considered. Performed system is run with the pressure control ventilation (PCV) which is one of the most used modes among other ventilation modes. With the simulation study, it can be said that the obtained datas from the outputs of fuzzy controller are suitable for patients having lung disease.
Abstract (Original Language): 
Bu çalıĢmada, bulanık mantık denetleyici ile yoğun bakım ünitelerinde kullanılan mekanik ventilatör cihazında tidal volüm, solunum sayısı ve hastaya ulaĢan gaz basıncı değerinin kontrolü gerçekleĢtirilmiĢtir. Bu sistemdeki amaç anestezistin iĢ yükünün azaltılmasıdır. Tidal volüm, solunum sayısı ve basınç değeri hesaplanır iken, referans basınç değeri (Pref) ile hastaya verilen gazın basınç değeri (Phasta) arasında ki hata Pe(k) ve bu hatanın değiĢim oranı Pe(k) dikkate alınmıĢtır. GerçekleĢtirilen sistem günümüzde çokça kullanılan ventilasyon modlarından olan Basınç Kontrollü Ventilasyon (PCV) modu ile çalıĢtırılmaktadır. Benzetim çalıĢması ile bulanık kontrolörün çıkıĢından elde edilen veriler akciğer rahatsızlığı olan hastalar için uygun olduğu söylenebilir.
98 - 105

REFERENCES

References: 

[1] Perel A, Stock Mc: Handbook of Mechanical Ventilatory
support. 1st Ed. Williams and Wilkins,Philadelphia,1992
[2] Kirby RR, Banner Mj, Downs JB(Eds), Ventilatory
Support 1st Ed Churchill Luingstone inc. Newyork 1990
[3]Smith RA, Respiratory Care MechanicalAnesthesia 2nd
Ed. Churchill Livingstone,New York ,1986
[4] Pillbeam SP „ Mechanical Ventilation: Physiological and
Clinical Application‟ 2nd Ed. St Louis,Mosby,1992
[5] A.Tzavaras, P.R.Weller, B.Spyropoulos, „A Neuro-
Fuzzy Controller for the Estimation of Tidal Volum and
Respiration Frequency Ventilator Settings for COPD
Patients Ventilated in Control Mode‟, Proceedings of the
Annual International Conference,of the IEEE,2007
[6] H.Zhu,K.Möller, „Ventilator Control Based on Fuzzy-
Neural Network Approach, Proceedings-30th International
Conference-IEEE/EMBS-2008
[7] D.S.Nelson,J.H.Strickland,T.C.Jannet „Simulation of
Fuzzy Control for Management of Respiratory Rate in
Assist Control Mechanical Ventilaton‟ Proceedings-19th
International Conference-IEEE/EMBS-1997
[8] C.S.Wang, D.Shaw, K.S.Jih „An Intelligent Control
System for Ventilators‟,Medical Engineering & Physics 20-
1998,534-542-Elseiver
[9] J.L.Larrabe, F.J.Alvarez ,E.G.Cuesta, A.V.Soller, L.F.
Alfonso „Development of a Time-Cycled Volume
Controlled Pressure-Limited Respirator and Lung
Mechanics System for Total Liquid Ventilator‟ IEEE
Transactions on Information Technology in Biomedicine
Vol.48 no.10-1134-1142-2001
[10]H.F.Kwok,D.A.Linkens,M.Mahfouf,G.H.Mills,C.L.Sim
pson,K.M.Goode „Fuzzy Logic Knowledge Elicitation for
Model Based Ventilator Management in the ICU‟
Proceedings-26th International Conference-IEEE/EMBS-
2002
[11] P.Cappa,S.A.Scuito „Experimental Analysis of the
Airway Circuit Effects on Breathing Pattern Generates by
Neonatal Pulmonary Ventilators‟ Proceedings-22nd
International Conference-IEEE/EMBS-2000
[12] H.Luepschen,L.Zhu,S.Leonhardt „Robust Closed Loop
Control of the Inspired Fraction of Oxygen for the Online
Assesment of Recruitment Maneuver‟ Proceedings-29th
International Conference-IEEE/EMBS-2007
[13] Güler.H, “Programlanabilir Lojik kontrolör ile Mekanik Ventilatör Tasarımı”, Fırat.Üniv.Fen Bilimleri Enst. Yük.Lis.Tezi 2007
[14] Elmas.Ç. “Bulanık Mantık Denetleyiciler”,Seçkin, Ankara,2003
[15] Siler,L.Ying,H. Fuzzy Control Theory.Elsevier. Fuzzy Set and System 33-1989
[16] Güler.H, ATA.F, “Calculation of Inspiration and Expiration Time with Fuzzy Logic Controller in Positive Pressure Ventilation" The 6th International Symposium On Intelligent & Manufacturing Systems. 14-17 October 2008-Sakarya