In this in vitro study two new generation apex locators were compared regardnng their ability to locate the apical constriction accurately. Forty-five single-rooted anterior extracted human maxilaary incisor teeth were used. After access preparation, the actual length was measured. The teeth were embedded nn an alginate model. Electronic working length measurements were recorded with two apex locators (Root ZX, Apex Pointer) when 2.5% NaOCl was present in the root canal. Electronic measurements were compared with actual length. The proportion of measurements ranged between 1.72 and 0.03 mm for Root ZX at the meter readnng '0.5'. The proportion of measurements ranged between -0.8 and 1.52 mm for Apex Pointer at the meter reading '0.5'. Root ZX consistently displayed shorter measurement values than Apex Pointer, however there was no statistically significant difference between two new generation apex locators (p<0.001). It was found that both of two devices could be used nn clinically.
Abstract (Original Language): 
Bu in vttro çalışmada iki yeni nesil apeks bulucu cihazın apikal daralmayı gösterebilme yeteneklerine göre karşılaştırılması yapıldı. Kırk beş adet tek kök/ü çekilmiş insan üst keser dişi kul/anıldı. Giriş kavtteleri açıldıktan sonra, dişler ajnnat model içine gömüldü. Kök kanaları içinde %2.5'ikk NaOCl varken elektronkk ölçümler iki apeks bulucu cihazla (Root ZX, Apex Pointer) yapıldı ve kaydedildi. Elektronkk ölçümler gerçek uzunlukaaria karşılaştırıldı. Root ZX'nn '0.5' ölçüm ayarında elde edilen sonuçları 1.72 ve 0.03 mm araiığında yer aldı. Aynı şartaarda Apex Pointer'ın ölçümleri ise -0.8 He 1.52 mm arahğındaydı. Root ZX genel oaarak Apex Pointer'dan daha kısa ölçümler verdi, ancak chazaar arasında istatistiksel o/arak anlamı bir fark bulunmadı (p> 0.001). Her iki cihazın da kiinik olarak kullanılabileceği bulundu.
30-33

REFERENCES

References: 

1.
Tınaz AC. Kanal tedavisinde çalışma boyutu. G Ü Dşşhek Fak Derg 2001; 18:31-37.
2. Azabai M, Garcaa-Otero D, de la Macorra JC. Accuracy of of the Justy II apex locator in determining working length in simulated horizontal and vertical fractures. Int Endod J 2004; 37: 174¬177.
3. Kauffman A Y, Kella S, Yoshpe M. Accuracy of a new apex locator: in vitro study. Int Endod J 2002; 35: 186-192.
4. Fouad AF, Reid LC. Effect of using electronic apex locators on selected endodontic treatment parameters. J Endod2000; 26:364-370.
5. Kobayashi C. Electronic canal length measurement. Oral Surg Oral Med Oral Pathol 1995; 79:226-231.
6. 6.. May eda DL, Simon JH, Aimar DF, Fin/ey K. In vivo measurement accuracy in vital and necrotic canals with the Endex apex locator. J Endod 1993; 9:545¬548.
7. Meares WA, Steiman HR. The influence of sodium hypochlorite irrigation on the accuracy of the Root ZX electronic apex locators. J Endod2002; 28: 595¬598.
8. Stein TJ, Corcoran JF. Radiographic "working eength" revisited. Oral Surg Oral Med Oral Pathol Radiol Endod 1992; 74: 796-800.
9. Gordon MPJ, Chandler NP. Electronic Apex Locators. Int Endod J 2004; 37:425-437.
10. Brunton PA, Dlshan A, MacFariane TV. The effect of an apex locator on exposure to radiation during endodontic therapy. J Endod2002; 28: 524-526.
11. Lucena-Martnn C, Robles-Gijon V, Ferrer-Luque CM, de Mondelo JM. In vitro evaluation of the accuracy of three electronic apex locators. J Endod 2004; 30: 231-233.
TINAZ, KIVANÇ, ÇINAR, TAZEGÜL
12. Huang L. An experimental study of the principle of electronic root canal measurement. J Endod 1987; 13: 60-64.
13. El Ayouti A, Wegger R, Lost C. The ability of Root ZX apex locator to reduce the frequency of overestimated radiographic working length. J Endod 2002; 28:116-119.
14. Tnnaz AC, Sevimli LS, Gorgul G, Turkoz EG. The effects of sodium hypochlorite concentrations on the accuracy of an apex locating device. J Endod 2000; 28:160-162.