Gama Kameraların Kalite Kontrol Testleri

Quality Control Tests of Gamma Cameras

Journal Name:

Publication Year:

Keywords (Original Language):

Abstract (2. Language): 
Thepurpose of nuclear medicine is diagnosis and treatment of diseases. Radionuclides are used in a routine imaging procedures of nuclear medicine. Radionuclides are combined with a chemical compound specified to related organ to form a radiopharmaceutical. These radiopharmaceuticals are introduced into patient's body by injection, swallowing or inhalation. The radioactive part of radiopharmaceutical emits gamma radiations and these gamma rays are detected by a gamma camera.Thus, the abnormal physiological and metabolic activities are detected in target organs. The gamma camera image quality is strictly dependent on regular quality control tests. There are a lot of numbers of factors that contribute to the image quality such as uniformity, resolution and collimation. For the performance checking of the gamma cameras, the National Electrical Manufacturers Association (NEMA) standards are applied. In this study, daily, weekly and monthly quality control tests of gamma camera installed in our department were performed. Performed quality control tests according to NEMA standards yielded acceptable results.
Abstract (Original Language): 
Nükleer tıbbın amacı hastalıkların teşhisi ve tedavisidir. Nükleer tıptaki rutin bir görüntüleme işleminde radyonüklidler kullanılır. Radyonüklidler organa özel bir kimyasal bileşik ile radyofarmasötik form oluşturulmak üzere birleştirilir. Bu radyofarmasötikler hasta vücuduna enjeksiyon, yutma veya soluma yolu ile verilir. Radyofarmasötiğin radyoaktif kısmı gama ışınları yayar ve bu ışınlar gama kamera tarafından dedekte edilir. Böylece anormal fizyolojik ve metabolik faaliyetler hedef organlarda tespit edilir. Gama kamera görüntü kalitesi düzenli kalite kontrol testlerine sıkı şekilde bağlıdır. Görüntü kalitesini etkileyen homojenite, rezolüsyon, kolimasyon gibi birçok faktör vardır. Gama kameraların performanslarını kontrol etmek için yapılan testlerde National Electrical Manufacturers Association (NEMA) standartları uygulanır. Bu çalışmada bölümümüzdeki gama kameranın günlük, haftalık ve aylık kalite kontrol testleri yapılmıştır. NEMA standartlarına göre yapılan kalite kontrol testleri uygun sonuçlar vermiştir.
75
84

REFERENCES

References: 

[1] Cherry S., Sorenson R., Phelps J. A., Michael E. 2012. Physics in Nuclear Medicine, Fourth Edition,
Elsevier
Saunders Publishing. [2] Standarts for performance measurements of scintillation cameras, Pub No NU-1, 2007, National
Electrical Manufacturers Association. NEMA, 2007. [3] Ellinor B. S., Plachcinska A., A Britten A., 2010. Routine quality control recommendations for
nuclear medicine instrumentation, European Journal of Nuclear Medicine and Molecular
Imaging, 37: 662-671.
[4]
Kılıç
Ş., Akdemir Ü.Ö., 2010. Gama kameraların kalite kontrol testleri, Turkish Journal of Clinical Laboratory, 1(1): 32-36.
83
D. Tunçman vd.
[5] Murphy P.H., 1987. Acceptance testing and quality control of gamma cameras, Including SPECT,
Journal of Nuclear Medicine, 28: 1221-1227. [6] Computer-Aided Scintillation Camera Acceptance Testing. A Task Group of the Nuclear Medicine
Committee American Association of Physicists in Medicine. AAPMReport No:9. [7] Rova A., Celler A., Hamarneh G., 2008. Development of NEMA-based Software for Gamma Camera
Quality Control. Journal of Digital Imaging, 21(2): 243- 255 [8] Gar-elnabi M. E. M., Ali W. M. , Ali Omer M. A., Edam G. A. 2015. Assessment of Gamma Camera
SPECT Resolution using Developed Quality Control Phantom, International Journal of Science
and Research .4(2): 2117-2121

Thank you for copying data from http://www.arastirmax.com