Journal Name:
- Advance Research in Pharmaceuticals and Biologicals
Key Words:
Author Name |
---|
Abstract (2. Language):
A HPTLC, UV spectrophotometric and ion pair extraction method were developed and validated for the estimation of
lumefantrine in tablets. The parameters linearity, precision, specificity, robustness, limit of detection and limit of
quantitation were studied according to “The International Conference on Harmonization Guidelines” for validation of
analytical procedures. The HPTLC method was performed using precoated silica gel plates GF60254 with mobile phase
ratio consisting of methanol : chloroform : ammonia (8:2:0.05 v/v/v). Densitometric scanning was performed at 268
nm. The Rf value was found to be 0.66. The linearity range were in the concentration range of 100- 500 ng/spot, 2-16
μg/ml, 5-25 μg/ml for HPTLC (method 1), UV spectrophotometric (method 2) and ion pair extraction method (method
3). The intraday precision and Interday precision were found to be less than 2 for all the three methods. The limit of
detection and limit of quantitation were found to be 25 ng/spot and 50 ng/spot for method 1, 0.5 μg/ml and 1 μg/ml
for method 2 and 2μg/ml and 4 μg/ml for method 3. Statistical analysis by student’s t-test showed no significant
difference between the results obtained by methods (p = 0.19497). The proposed methods are highly sensitive, precise
and accurate and can be used for the routine quantitation of lumefantrine in tablets.
Bookmark/Search this post with
FULL TEXT (PDF):
- 3
296-299